Nebulizer
Nebulizer
Jump to navigation
Jump to search
Nebulizer | |
---|---|
A hospital nebulizer setup | |
Specialty | pulmonology |
[edit on Wikidata] |
In medicine, a nebulizer[1] or nebuliser[2] (see spelling differences) is a drug delivery device used to administer medication in the form of a mist inhaled into the lungs. Nebulizers are commonly used for the treatment of asthma, cystic fibrosis, COPD and other respiratory diseases or disorders.
Analytical nebulizers are another form of nebulizer that is used primarily in laboratory settings for elemental analysis.
Nebulizers use oxygen, compressed air or ultrasonic power to break up solutions and suspensions into small aerosol droplets that can be directly inhaled from the mouthpiece of the device. An aerosol is a mixture of gas and solid or liquid particles.
Contents
1 Medical uses
1.1 Guidelines
1.2 Effectiveness
2 Aerosol deposition
3 Types of nebulizers
3.1 Mechanical
3.1.1 Soft mist inhaler
3.2 Electrical
3.2.1 Jet nebulizer
3.2.2 Ultrasonic wave nebulizer
3.2.3 Vibrating mesh technology
4 Use and attachments
5 History
6 See also
7 References
Medical uses[edit]
Guidelines[edit]
Various asthma guidelines, such as the Global Initiative for Asthma Guidelines [GINA], the British Guidelines on the management of Asthma, The Canadian Pediatric Asthma Consensus Guidelines, and United States Guidelines for Diagnosis and Treatment of Asthma each recommend metered dose inhalers in place of nebulizer-delivered therapies.[3]
The European Respiratory Society acknowledge that although nebulizers are used in hospitals and at home they suggest much of this use may not be evidence-based.[4]
Effectiveness[edit]
Recent evidence shows that nebulizers are no more effective than metered-dose inhalers (MDIs) with spacers.[5] An MDI with a spacer may offer advantages to children who have acute asthma.[3][6][5] Those findings refer specifically to the treatment of asthma and not to the efficacy of nebulisers generally, as for COPD for example.[5] For COPD, especially when assessing exacerbations or lung attacks, there is no evidence to indicate that MDI (with a spacer) delivered medicine is more effective than administration of the same medicine with a nebulizer.[7]
The European Respiratory Society highlighted a risk relating to droplet size reproducibility caused by selling nebulizer devices separately from nebulized solution. They found this practice could vary droplet size 10-fold or more by changing from an inefficient nebulizer system to a highly efficient one.[4][5]
Two advantages attributed to nebulizers, compared to MDIs with spacers (inhalers), were their ability to deliver larger dosages at a faster rate, especially in acute asthma; however, recent data suggests actual lung deposition rates are the same. In addition, another trial found that a MDI (with spacer) had a lower required dose for clinical result compared to a nebulizer (see Clark, et al. other references).[3]
Aerosol deposition[edit]
The lung deposition characteristics and efficacy of an aerosol depend largely on the particle or droplet size. Generally, the smaller the particle the greater its chance of peripheral penetration and retention. However, for very fine particles below 0.5 µm in diameter there is a chance of avoiding deposition altogether and being exhaled. In 1966 the Task Group on Lung Dynamics, concerned mainly with the hazards of inhalation of environmental toxins, proposed a model for deposition of particles in the lung. This suggested that particles of more than 10 µm in diameter are most likely to deposit in the mouth and throat, for those of 5–10 µm diameter a transition from mouth to airway deposition occurs, and particles smaller than 5 µm in diameter deposit more frequently in the lower airways and are appropriate for pharmaceutical aerosols.[8]
Types of nebulizers[edit]
Mechanical[edit]
Soft mist inhaler[edit]
The medical company Boehringer Ingelheim also invented a new device named Respimat Soft Mist Inhaler in 1997. This new technology provides a metered dose to the user, as the liquid bottom of the inhaler is rotated clockwise 180 degrees by hand, adding a build up tension into a spring around the flexible liquid container. When the user activates the bottom of the inhaler, the energy from the spring is released and imposes pressure on the flexible liquid container, causing liquid to spray out of 2 nozzles, thus forming a soft mist to be inhaled. The device features no gas propellant and no need for battery/power to operate. The average droplet size in the mist was measured to a somewhat disappointing 5.8 micrometers, which could indicate some potential efficiency problems for the inhaled medicine to reach the lungs. Subsequent trials have proven this was not the case. Due to the very low velocity of the mist, the Soft Mist Inhaler in fact has a higher efficiency compared to a conventional pMDI.[9] In 2000, arguments were launched towards the European Respiratory Society (ERS) to clarify/expand their definition of a nebulizer, as the new Soft Mist Inhaler in technical terms both could be classified as a "hand driven nebulizer" and a "hand driven pMDI".[10]
Electrical[edit]
Jet nebulizer[edit]
The most commonly used nebulizers are jet nebulizers, which are also called "atomizers".[11] Jet nebulizers are connected by tubing to a compressor, that causes compressed air or oxygen to flow at high velocity through a liquid medicine to turn it into an aerosol, which is then inhaled by the patient. Currently there seems to be a tendency among physicians to prefer prescription of a pressurized Metered Dose Inhaler (pMDI) for their patients, instead of a jet nebulizer that generates a lot more noise (often 60 dB during use) and is less portable due to a greater weight. However, jet nebulizers are commonly used for patients in hospitals who have difficulty using inhalers, such as in serious cases of respiratory disease, or severe asthma attacks.[12] The main advantage of the jet nebulizer is related to its low operational cost. If the patient needs to inhale medicine on a daily basis the use of a pMDI can be rather expensive. Today several manufacturers have also managed to lower the weight of the jet nebulizer down to 635 grams (22.4 oz), and thereby started to label it as a portable device. Compared to all the competing inhalers and nebulizers, the noise and heavy weight is however still the biggest draw back of the jet nebulizer. Trade names for jet nebulizers include Maxin.[13]
Ultrasonic wave nebulizer[edit]
Ultrasonic wave nebulizers were invented in 1965[14] as a new type of portable nebulizer. The technology inside an ultrasonic wave nebulizer is to have an electronic oscillator generate a high frequency ultrasonic wave, which causes the mechanical vibration of a piezoelectric element. This vibrating element is in contact with a liquid reservoir and its high frequency vibration is sufficient to produce a vapor mist.[15] As they create aerosols from ultrasonic vibration instead of using a heavy air compressor, they only have a weight around 170 grams (6.0 oz). Another advantage is that the ultrasonic vibration is almost silent. Examples of these more modern type of nebulizers are: Omron NE-U17 and Beurer Nebulizer IH30.[16]
Vibrating mesh technology[edit]
A new significant innovation was made in the nebulizer market around 2005, with creation of the ultrasonic Vibrating Mesh Technology (VMT). With this technology a mesh/membrane with 1000-7000 laser drilled holes vibrates at the top of the liquid reservoir, and thereby pressures out a mist of very fine droplets through the holes. This technology is more efficient than having a vibrating piezoelectric element at the bottom of the liquid reservoir, and thereby shorter treatment times are also achieved. The old problems found with the ultrasonic wave nebulizer, having too much liquid waste and undesired heating of the medical liquid, have also been solved by the new vibrating mesh nebulizers. Available VMT nebulizers include: Pari eFlow,[17]Respironics i-Neb,[18] Beurer Nebulizer IH50,[19] and Aerogen Aeroneb.[20] As the price of the ultrasonic VMT nebulizers is higher than models using previous technologies, most manufacturers continue to also sell the classic jet nebulizers.
Use and attachments[edit]
Nebulizers accept their medicine in the form of a liquid solution, which is often loaded into the device upon use. Corticosteroids and bronchodilators such as salbutamol (albuterol USAN) are often used, and sometimes in combination with ipratropium. The reason these pharmaceuticals are inhaled instead of ingested is in order to target their effect to the respiratory tract, which speeds onset of action of the medicine and reduces side effects, compared to other alternative intake routes.[12]
Usually, the aerosolized medicine is inhaled through a tube-like mouthpiece, similar to that of an inhaler. The mouthpiece, however, is sometimes replaced with a face mask, similar to that used for inhaled anesthesia, for ease of use with young children or the elderly. Pediatric masks are often shaped like animals such as fish, dogs or dragons to make children less resistant to nebulizer treatments. Many nebulizer manufacturers also offer pacifier attachments for infants and toddlers. But mouthpieces are preferable if patients are able to use them since face-masks result in reduced lung delivery because of aerosol losses in the nose.[11]
After use with corticosteroid, it is theoretically possible for patients to develop a yeast infection in the mouth (thrush) or hoarseness of voice (dysphonia), although these conditions are clinically very rare. To avoid these adverse effects, some clinicians suggest that the person who used the nebulizer should rinse his or her mouth. This is not true for bronchodilators; however, patients may still wish to rinse their mouths due to the unpleasant taste of some bronchodilating drugs.
History[edit]
The first "powered" or pressurized inhaler was invented in France by Sales-Girons in 1858.[21] This device used pressure to atomize the liquid medication. The pump handle is operated like a bicycle pump. When the pump is pulled up, it draws liquid from the reservoir, and upon the force of the user's hand, the liquid is pressurized through an atomizer, to be sprayed out for inhalation near the user's mouth.[22]
In 1864, the first steam-driven nebulizer was invented in Germany. This inhaler, known as "Siegle’s steam spray inhaler", used the Venturi principle to atomize liquid medication, and this was the very beginning of nebulizer therapy. The importance of droplet size was not yet understood, so the efficacy of this first device was unfortunately mediocre for many of the medical compounds. The Siegle steam spray inhaler consisted of a spirit burner, which boiled water in the reservoir into steam that could then flow across the top and into a tube suspended in the pharmaceutical solution. The passage of steam drew the medicine into the vapor, and the patient inhaled this vapor through a mouthpiece made of glass.[23]
The first electrical nebulizer was invented in the 1930s and called a Pneumostat. With this device, a medical liquid (typically epinephrine chloride, used as a bronchial muscle relaxant to reverse constriction) was made aerosol by the power from an electrical compressor.[24] As an alternative to the expensive electrical nebulizer, many people in the 1930s continued to use the much more simple and cheap hand-driven nebulizer, known as the Parke-Davis Glaseptic.[25]
In 1956, a technology competing against the nebulizer was launched by Riker Laboratories (3M), in the form of pressurized metered-dose inhalers, with Medihaler-iso (isoprenaline) and Medihaler-epi (epinephrine) as the two first products.[26] In these devices, the drug is cold-fill and delivered in exact doses through some special metering valves, driven by a gas propellant technology (i.e. Freon or a less environmentally damaging HFA).[21]
In 1964, a new type of electronic nebulizer was introduced: the "ultrasonic wave nebulizer".[27] Today the nebulizing technology is not only used for medical purposes. Ultrasonic wave nebulizers are also used in humidifiers, to spray out water aerosols to moisten dry air in buildings.[15]
Some of the first models of electronic cigarettes featured an ultrasonic wave nebulizer (having a piezoelectric element vibrating and creating high-frequency ultrasound waves, to cause vibration and atomization of liquid nicotine) in combination with a vapouriser (built as a spray nozzle with an electric heating element).[28] The most common type of electronic cigarettes currently sold, however, omit the ultrasonic wave nebulizer, as it was not found to be efficient enough for this kind of device. Instead, the electronic cigarettes now use an electric vaporizer, either in direct contact with the absorbent material in the "impregnated atomizer," or in combination with the nebulization technology related to a "spraying jet atomizer" (in the form of liquid droplets being out-sprayed by a high-speed air stream, that passes through some small venturi injection channels, drilled in a material absorbed with nicotine liquid).[29]
See also[edit]
- Inhaler
- Humidifier
- Vaporizer
- List of medical inhalants
- Spray bottle
References[edit]
^ Medical Dictionary. "Nebulizer". Retrieved 2010-11-01..mw-parser-output cite.citationfont-style:inherit.mw-parser-output .citation qquotes:"""""""'""'".mw-parser-output .citation .cs1-lock-free abackground:url("//upload.wikimedia.org/wikipedia/commons/thumb/6/65/Lock-green.svg/9px-Lock-green.svg.png")no-repeat;background-position:right .1em center.mw-parser-output .citation .cs1-lock-limited a,.mw-parser-output .citation .cs1-lock-registration abackground:url("//upload.wikimedia.org/wikipedia/commons/thumb/d/d6/Lock-gray-alt-2.svg/9px-Lock-gray-alt-2.svg.png")no-repeat;background-position:right .1em center.mw-parser-output .citation .cs1-lock-subscription abackground:url("//upload.wikimedia.org/wikipedia/commons/thumb/a/aa/Lock-red-alt-2.svg/9px-Lock-red-alt-2.svg.png")no-repeat;background-position:right .1em center.mw-parser-output .cs1-subscription,.mw-parser-output .cs1-registrationcolor:#555.mw-parser-output .cs1-subscription span,.mw-parser-output .cs1-registration spanborder-bottom:1px dotted;cursor:help.mw-parser-output .cs1-ws-icon abackground:url("//upload.wikimedia.org/wikipedia/commons/thumb/4/4c/Wikisource-logo.svg/12px-Wikisource-logo.svg.png")no-repeat;background-position:right .1em center.mw-parser-output code.cs1-codecolor:inherit;background:inherit;border:inherit;padding:inherit.mw-parser-output .cs1-hidden-errordisplay:none;font-size:100%.mw-parser-output .cs1-visible-errorfont-size:100%.mw-parser-output .cs1-maintdisplay:none;color:#33aa33;margin-left:0.3em.mw-parser-output .cs1-subscription,.mw-parser-output .cs1-registration,.mw-parser-output .cs1-formatfont-size:95%.mw-parser-output .cs1-kern-left,.mw-parser-output .cs1-kern-wl-leftpadding-left:0.2em.mw-parser-output .cs1-kern-right,.mw-parser-output .cs1-kern-wl-rightpadding-right:0.2em
^ British Spelling of Nebulizer Medical Dictionary. "Definition". Archived from the original on 2010-07-01. Retrieved 2010-11-01.
^ abc Clark NM, Houle C, Partridge MR, Leo HL, Paton JY (2010). "The puzzle of continued use of nebulized therapy by those with asthma". Chron Respir Dis. 7 (1): 3–7. doi:10.1177/1479972309357496. PMID 20103617.
^ ab Boe J, Dennis JH, O'Driscoll BR, et al. (July 2001). "European Respiratory Society Guidelines on the use of nebulizers". Eur. Respir. J. 18 (1): 228–42. doi:10.1183/09031936.01.00220001. PMID 11510796.
^ abcd Cates, Christopher J.; Welsh, Emma J.; Rowe, Brian H. (2013-09-13). "Holding chambers (spacers) versus nebulisers for beta-agonist treatment of acute asthma". The Cochrane Database of Systematic Reviews (9): CD000052. doi:10.1002/14651858.CD000052.pub3. ISSN 1469-493X. PMID 24037768.
^ Epling J, Chang MH (January 2003). "Are metered-dose inhalers with holding chambers better than nebulizers for treating acute asthma?". Am Fam Physician. 67 (1): 62–4. PMID 12537167.
^ van Geffen, Wouter H.; Douma, W. R.; Slebos, Dirk Jan; Kerstjens, Huib A. M. (2016-08-29). "Bronchodilators delivered by nebuliser versus pMDI with spacer or DPI for exacerbations of COPD" (PDF). The Cochrane Database of Systematic Reviews (8): CD011826. doi:10.1002/14651858.CD011826.pub2. ISSN 1469-493X. PMID 27569680.
^ The science of nebulised drug delivery, p6
^ Boehringer Ingelheim (2003). "How it works: Respimat Soft Mist Inhaler". Archived from the original on 2007-05-27. Retrieved 2005-08-16.
^ Denyer J, et al. (2000). "New liquid drug aerosol devices for inhalation therapy". Eur Respir Rev. 10: 187–191.
^ ab Finlay, W.H. (2001). The Mechanics of Inhaled Pharmaceutical Aerosols: An Introduction. Academic Press.
^ ab Hickey, A.J. (2004). Pharmaceutical Inhalation Aerosol Technology (2nd ed.). New York: Marcel Dekker.
^ J. Jendle; B. E. Karlberg; J. Persliden; L. Franzen; M. Jr Arborelius (Fall 1995). "Delivery and retention of an insulin aerosol produced by a new jet nebulizer". Journal of Aerosol Medicine. 8 (3): 243–254. doi:10.1089/jam.1995.8.243. PMID 10155650.
^ US patent 3243122, Snaper, Alvin A., "Ultrasonic Spray Apparatus", published 29 March 1966
^ ab BOGA Gmbh. "Operating principle of ultrasonic humidifier". Archived from the original on 2010-11-14. Retrieved 2010-04-05.
^ Knoch, M.; Finlay, W.H. (2002). "Ch. 71 Nebuliser Technologies". In Rathbone, Hadgraft, Roberts. Modified-Release Drug Delivery Technology. Marcel Dekker. pp. 849–856.CS1 maint: Uses editors parameter (link)
^ PARI Pharma (2008). "Leading aerosol therapies worldwide, delivery with eFlow". Archived from the original on 2010-03-28. Retrieved 2010-04-09.
^ Philips Respironics (2010). "Active Aerosol Delivery, The I-neb and Vibrating Mesh Technology". Archived from the original on 2010-08-04. Retrieved 2010-04-09.
^ Beurer (2015). "Product details of IH50 nebulizer, with a vibrating membrane". Retrieved 2015-04-21.
^ Aerogen (2009). "Micropump nebulizers, Aeroneb, Vibrating Mesh Technology". Archived from the original on 2010-02-03. Retrieved 2010-04-09.
^ ab Sanders M (April 2007). "Inhalation therapy: an historical review" (PDF). Prim Care Respir J. 16 (2): 71–81. doi:10.3132/pcrj.2007.00017. PMID 17356785.
^ Inhalatorium. "Pressurized inhaler invented by Sales-Girons". Archived from the original on 2013-01-03. Retrieved 2010-04-05.
^ Inhalatorium. "Siegle's steam spray inhaler". Archived from the original on 2004-08-26. Retrieved 2010-04-05.
^ Inhalatorium. "First electrical nebulizer (Pneumostat)". Archived from the original on 2005-02-17. Retrieved 2010-04-05.
^ Inhalatorium. "The hand driven nebulizer "Parke-Davis Glaseptic". Archived from the original on 2004-09-06. Retrieved 2010-04-05.
^ Riker Laboratories (1960-03-16) [1956-03-21]. "Self-propelling pharmaceutical compositions (for a pMDI)". GB patent. Retrieved 24 Jan 2016.
^ Devilbiss Co. (1967-05-17) [1964-02-10]. "Method and apparatus for producing aerosols (ultrasonic nebulizer)". GB patent. Retrieved 24 Jan 2016.
^ Hon Lik (2004-04-14). "An aerosol electronic cigarette". CN patent. Retrieved 2006-12-27.
^ Hon Lik (2006-05-16). "Emulation aerosol sucker". CN patent. Retrieved 2009-02-11.
Categories:
- Aerosols
- Respiratory therapy
- Drug delivery devices
- Medical equipment
- Dosage forms
(window.RLQ=window.RLQ||).push(function()mw.config.set("wgPageParseReport":"limitreport":"cputime":"0.572","walltime":"0.699","ppvisitednodes":"value":2002,"limit":1000000,"ppgeneratednodes":"value":0,"limit":1500000,"postexpandincludesize":"value":104366,"limit":2097152,"templateargumentsize":"value":1011,"limit":2097152,"expansiondepth":"value":10,"limit":40,"expensivefunctioncount":"value":2,"limit":500,"unstrip-depth":"value":1,"limit":20,"unstrip-size":"value":86015,"limit":5000000,"entityaccesscount":"value":3,"limit":400,"timingprofile":["100.00% 567.870 1 -total"," 63.54% 360.850 1 Template:Reflist"," 30.75% 174.635 16 Template:Cite_web"," 18.48% 104.965 1 Template:Infobox_medical_equipment"," 17.13% 97.283 1 Template:Infobox"," 16.79% 95.331 8 Template:Cite_journal"," 9.05% 51.403 6 Template:Navbox"," 6.50% 36.918 2 Template:Convert"," 6.39% 36.298 1 Template:Dosage_forms"," 3.98% 22.604 1 Template:Cite_patent"],"scribunto":"limitreport-timeusage":"value":"0.322","limit":"10.000","limitreport-memusage":"value":7301356,"limit":52428800,"cachereport":"origin":"mw1265","timestamp":"20190304011914","ttl":2592000,"transientcontent":false);mw.config.set("wgBackendResponseTime":855,"wgHostname":"mw1265"););